Article Summary
唐云波,郑林丰,欧阳欣,许静波,王治愚.256排Revolution多参数调节下低剂量心脏冠脉成像应用[J].现代生物医学进展英文版,2018,(10):1922-1926.
256排Revolution多参数调节下低剂量心脏冠脉成像应用
Application of 256-row Revolution CT with Multi-parameter Regulation for Low-dose Coronary Computed Tomographic Angiography
Received:December 27, 2017  Revised:January 22, 2018
DOI:10.13241/j.cnki.pmb.2018.10.023
中文关键词: 心脏冠脉成像  多参数调节  自动毫安  自动千伏  图像迭代重建
英文关键词: Coronary computed tomographic angiography  Multi-parameter Regulation  Automatic milliampere  Automatic kV  Im- age iterative reconstruction
基金项目:上海市浦江人才计划项目(17PJ1408000) ;上海市自然科学基金资助项目(17ZR1422500);上海交通大学医工(理)交叉基金资助项目(YG2016MS26)
Author NameAffiliationE-mail
唐云波 湖南省湘潭市中心医院放射科 湖南 湘潭 411100 tangyunbotyb@163.com 
郑林丰 上海交通大学附属第一人民医院放射科 上海 200080  
欧阳欣 湖南省湘潭市中心医院放射科 湖南 湘潭 411100  
许静波 上海交通大学附属第一人民医院放射科 上海 200080  
王治愚 上海交通大学附属第一人民医院放射科 上海 200080  
Hits: 488
Download times: 318
中文摘要:
      摘要 目的:研究256排Revolution多参数调节下低剂量扫描心脏冠脉成像(Coronary computed tomographic angiography,CCTA)的应用。方法:回顾性分析2016年3月15日至5月15 日间在我院256排CT行CCTA的患者,分为心率稳定试验组(n=54例)和心率不稳定试验组(n=41例),同时选取在我院64排VCT行CCTA的患者为对照组(n=116例),对三组CCTA图像质量行主观和客观评估,比较有效辐射剂量的差异。结果:256排CT检查成功率和图像质量的主观评价均优于对照组(P<0.05);信噪比(SNR)、对比噪声比(CNR)与对照组差异无统计学意义(P>0.05);但有效辐射剂量明显降低(P<0.05)。结论:256排CT多参数调节下检查冠脉在保证图像质量下可有效降低辐射剂量。
英文摘要:
      ABSTRACT Objective: To study the feasibility of 256-row revolution computed tomography (CT) with optimal multi-parameter se- lection for low-dose coronary computed tomographic angiography (CCTA). Methods: This retrospective study enrolled 95 cases under- going CCTA with 256-row CT in our hospital from March 15,2016 to May 15, 2016 and these patients was divided into two groups: sta- ble heart rate group (n=54) and unstable heart rate group (n=41). Additionally, we selected 116 cases whom underwent CCTA examina- tion with 64-slice VCT as control group (n=116). Then, we evaluated the image quality of subjective and objective score and compared effective radiation dose in these three groups. Results: For CCTA with 256-row CT, both the success rate and subjective evaluation of im- age quality were superior to conventional 64-slice VCT (control group, P<0.05). The signal-to-noise ratio (SNR) and contrast-to-noise ra- tio (CNR) were not statistically significant between the three group (P>0.05). However, the radiation dose was significantly reduced than the control group (P<0.05). Conclusion: With optimal multi-parameter regulation, 256-row revolution CT can effectively reduce the radi- ation dose while preserve diagnostic image quality for CCTA.
View Full Text   View/Add Comment  Download reader
Close