杨玉刚 尚革 许林 齐洪志 郝洁.宫颈癌调强放射治疗的剂量验证分析[J].现代生物医学进展英文版,2014,14(23):4526-4528. |
宫颈癌调强放射治疗的剂量验证分析 |
Verification and Analysis of the Dose of Intensity Modulation Radiated Therapy(IMRT)on Cervical Cancer |
|
DOI: |
中文关键词: 宫颈癌 调强放疗 剂量验证 γ通过率 Xio Eclipse |
英文关键词: Cervical cancer IMRT(Intensity Modulation Radiated Therapy) Dose verification γpass rate Xio Eclipse |
基金项目:乌鲁木齐市科学技术计划项目,感染与肿瘤重点实验室(H111313001) |
|
Hits: 755 |
Download times: 980 |
中文摘要: |
目的:通过总结比较Eclipse 和Xio 两个治疗计划系统对宫颈癌调强放疗剂量验证的结果,分析评价宫颈癌调强放疗在临床
应用上剂量的准确性。方法:分别用固体水模和二维电离室矩阵Mapcheck、FC65-G指形电离室在定位螺旋CT上进行扫描,断层
数据通过网络传输到Eclipse 和Xio 两个治疗计划系统中建立验证模体。选取2013 年1月至5 月在本院24 例宫颈癌调强治疗患
者,在两个TPS(Treatment Planning Systems)上将宫颈癌调强放疗计划移植到模体上,用IBA公司的DOSE1、FC65-G 指形电离室
模体和Nuclear 公司的二维电离室矩阵Mapcheck 模体在加速器下进行点剂量和面剂量的采集。结果:Eclipse 系统的12 位患者
随机选取24 个照射野3 mm 3%面剂量的酌通过率均值为98.8%,Xio 系统的12 位患者随机选取24 个照射野3mm3%面剂量γ
通过率均值为97.7%,二者比较差别有统计学意义(P<0.01)。Eclipse 系统的12 位患者84 个照射野点剂量的误差的均值为
1.2417%,Xio 系统的12 位患者84 个照射野点剂量的误差均值为2.4892%,二者比较差别具有统计学意义(P<0.01)。结论:Eclipse
和Xio 两个治疗计划系统都可以应用于宫颈癌临床调强放疗,但Eclipse 系统的准确性优于Xio 系统。 |
英文摘要: |
Objective:To verify and analyze the dose of intensity modulation radiated therapy (IMRT) on cervical cancer through
comparing the Eclipse and Xio systems and evaluate the accuracy of dose of cervical cancer in the clinical application of IMRT.Methods:Solid water phantom and two-dimensional ionization chamber matrix Mapcheck, FC65-G ionization chamber were used in the
positioning finger on spiral CT scans, fault data through the network to the Eclipse and Xio treatment planning system to create
verification phantom. 24 cases of cervical cancer admitted between January and May 2013 in the hospital were selected and treated with
IMRT at two TPS (Treatment Planning Systems), cervical cancer IMRT plans will be transplanted to the verification phantom, and with
the IBA's DOSE1, FC65-G ionization finger-chamber and Nuclear company's two-dimensional matrix Mapcheck phantom ionization
chamber with verification phantomunder the accelerator to get point dose and the surface dose.Results:Eclipse system, 12 patients were
randomly selected 24 radiation fields 3 mm3% surface dose of γrate value through 98.8%, Xio system, 12 patients were randomly
selected 24 radiation fields 3 mm3%surface dose γrate is 97.7%(P <0.01). The mean error of the 12 patients 84 point dose of irradiation
fields in Eclipse system was 1.2417%, the mean error of 12 patients 84 point dose irradiation fields in Xio system was 2.4892%(P<0.01).Conclusion:Two treatment planning systems could be applied to cervical cancer IMRT, the accuracy of Eclipse system was superior to
that of Xio system . |
View Full Text
View/Add Comment Download reader |
Close |