文章摘要
高伟 刘杨月 高晶 邱赫 方芳.超低剂量CT引导下胸部穿刺活检的应用[J].,2016,16(1):135-137
超低剂量CT引导下胸部穿刺活检的应用
Application of Low-dose CT-Guided Needle Biopsy of Chest
  
DOI:
中文关键词: 活组织检查  辐射剂量  CT
英文关键词: Biopsy  Needle  Radiation dose  CT
基金项目:黑龙江省卫生厅资助课题(2013-155)
作者单位
高伟 刘杨月 高晶 邱赫 方芳 哈尔滨医科大学附属第四医院医学工程科黑龙江中医药大学附属第二医院针灸科 
摘要点击次数: 779
全文下载次数: 0
中文摘要:
      目的:研究超低剂量扫描在CT 引导下肺部穿刺活检中的应用。方法:2009年1 月至2014 年12 月来本院200例行胸部CT 引导下穿刺的患者分为常规剂量组(130 kV,90 mAs)、超低剂量组(100 kV,7.5 mAs),各组100 例,并进行穿刺。根据CT 机自动记 录的受检者剂量长度乘积(dose length product, DLP)计算X 线有效辐射剂量(effective dose, ED)。比较两组在辐射剂量、图片质 量、穿刺成功率的差异。结果:超低剂量组和常规剂量组DLP 分别为(25.79 ± 1.13) mGy·cm 和(256.14 ± 4.48) mGy·cm(P< 0.05),ED 分别为(0.36 ± 0.02) mSV 和(3.59 ± 0.06) mSv(P<0.05)。图片质量均为A级。超低剂量组和常规剂量组穿刺成功率分别 为88 %(88/100)为92 %(92/100),差异无统计学意义(x2=0.89, P>0.05)。结论:超低剂量CT 引导下胸部穿刺,可降低受检者的X 线有效辐射剂量,图像质量可达到临床CT 引导下胸部穿刺的要求,值得提倡和推广。
英文摘要:
      Objective:To study application of low dose CT-guided puncture biopsy of lung.Methods:From January 2009 to December 2014, 200 consecutive patients prepared for lung biopsy were divided into standard dose group and low-dose group. Of these patients, 100 patients underwent percutaneous needle lung biopsy under standard dose CT scan (130 kV, 90 mAs), while the other 100 patients underwent low-dose CT scan (100 kV, 7.5 mAs) and puncture. The effective dose (ED) was calculated by dose length product (DLP) automatically recorded by the CT machine. And the statistical analyses on ED and DLP from two groups were performed. Data from two groups were analyzed for radiation dose, image quality and puncture success rate.Results:The mean DLP and ED were (25.79± 1.13) mGy·cm and (0.36± 0.02) mSv in low-dose group and (256.14± 4.48) mGy·cm and(3.59± 0.06) mSv in the standard- dose group respectively, both parameters indicating significant difference between two groups respectively (P<0.05). The image quality of the two groups were both level A. Also, there was no statistical difference on success puncture rate (x2=0.89, P>0.05) in low-dose group (88 %, 88/100) compared to the standard-dose group (92 %, 92/100).Conclusion:Low-dose chest CT-guided biopsy can significantly reduce the radiation dose, while the image quality meet requirements of clinical chest CT-guided biopsy.
查看全文   查看/发表评论  下载PDF阅读器
关闭